PROCEDURES
FOR DOA BOTTLEMULTIPLEX |
| |
|
|
BENEFITS
• Simple and rapid
• No special equipment is needed
• Results are easy to interpret
• No sample preparation
• Results in 5 minutes |
|
i. Remove
cap and fill in urine to above the minimum
level.

|
| |
|
|
PERFORMANCE
CHARACTERISTICS
Method : Immunochromatography
Detection : Opiate : 300 ng/ml or
> Marijuana (THC) : 50 ng/ml or >
Methamphetamine :1000 ng/ml
or > Amphetamine :1000 ng/ml
or > Specimen :
Urine
Assay time : 5 minutes
Reproducibility : Very high
Precision : Very good |
|
ii.
Screw the cap properly. Label the cup accordingly.
Twist the lock clockwise
until it
tightens (once only).

|
| |
|
|
APPLICATION
Pre-employment or applicant testing.
Scheduled testing.
Random testing.
Post-accident or for-cause testing.
Treatment follow-up testing.
ORDERING
INFORMATION
MBDr/DOA05 |
|
iii.
Immediately tilt the bottle with the cassette
facing towards you, then place the
bottle on a flat surface. Read results
in 5 minutes.

|
INTERPRETATION
OF RESULTS
1. POSITIVE
: The appearance of a reddish-purple control line
and no distinct line on a specific drug position
indicates the test result is positive for that
drug (i.e. the specimens contains the drug at a
concentration above the cut off level). A positive
test result does not provide any indication of
the level of intoxication or urinary concentrate
of the drug in the sample, it only indicates
the sample contains drug above the cut off level
in qualitative terms.
2. NEGATIVE : The appearance of a reddish-purple
control line and a line on a specific drug position
indicates a negative test result i.e. no drug
above the cut off level has been detected. The
color intensities of the control line and a specific
drug line may not be equal. Any faint line at
the specific drug position, visible in 5 minutes,
should indicate the absence of drug in the sample.
It only indicates that the sample does not contain
drug above the cut off level in qualitative terms.
3. INVALID : A distinct control line should always
appear. The test is invalid if no control line
is seen. Such a test should be repeated with
a new DOA BOTTLE MULTIPLEX (OPI/AMP/MET/THC)
test device.
|
|
iv.

|
|
|
|
 |
|
|